Features
1.Two operation modes fluorescence intensity and luminous intensity.
2.365nm exciting wavelength,Raman peak of water S/N ≥150.
3.10 stages emission spectrum scanning.
4. Support off-line mode and on-line mode.
Model |
MF-FS96 |
MF-FS93 |
Light source |
Hamamatsu 150W Xenon lamp |
LED |
Exciting optical filters |
Interference optical filter
a) Standard set is equipped with an interference optical filter of
b) An optional interference optical filer of 25mm diameter of |
Interference optical filter (10nm bandwidth) |
Emission Monochromator |
C-T diffraction grating(Em 200-900nm,bandwidth 10nm) |
360-650nm (C-T diffraction grating) |
Wavelength Accuracy |
±1nm |
±2nm |
Wavelength Repeatability |
≤0.5nm |
≤1nm |
Sensitivity |
Raman peak of water in 1 cm quartz fluorescence cuvette with S/N≥150 |
1×10-9 g/ml (Standard aqua of the sulfuric acid Quinine, 12nm bandwidth) |
Linear measurement(r) |
≥0.995 |
≥0.995 |
Stability |
better than 1.5%/10min |
better than 3%/10min |
Response Time |
(0.1-4)s 6 stages adjustable |
(0.1-4)s |
Fluorescence display value |
0.00-600.00 |
0.00-250.00 |
Power Supply |
AC110/220V±10%, 60/50Hz |
|
Packing Size(mm) |
550*510*365 |
450* 420 *280 |
Gross Weight(kg) |
14 |
9 |
Standard Accessories
|
365nm filter(Preassembled), 1 pc FS96 software package, 1 pc USB interface, 1 pc Quartz fluorescence sample cell10mm, 1 pair |
RS232 serial port ,Glass fluorescence cuvette 10mm 1 pair |
Optional Accessories |
200-700nm interference optical filter (φ25mm) Quartz fluorescence sample cell10mm Personal computer Dedicated serial interface printer |
Data processing software package (pack, for PC), Printer(Serial printer/Jet printer) |
Medfutureusa